TB 9-6625-2009-24
(7) Disconnect cable from multimeter and connect to TI INPUT (front panel).
(8) Press and release DISPLAY CLEAR WRITE pushbutton while adjusting
SWEEP MODE MANUAL VERNIER control for a centered narrow spike display.
(9) Adjust TRACKING OSC LEVEL control (rear panel) for a full scale 2 V display.
(10) Adjust RESOLUTION BANDWIDTH control to 3 Hz. TI display will drop no
more than 5 divisions.
(11) Adjust FREQ SPAN/DIV control to 0.1 kHz and adjust SWEEP MODE
MANUAL VERNIER control for a 1 kHz display on far right graticule line.
(12) Connect audio analyzer to TI TRACKING OSC OUT (rear panel).
(13) Rotate TRACKING OSC LEVEL control (rear panel) fully cw. Audio analyzer
indication will be 1 percent or less.
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.