SECTION III. CALIBRATION PROCESS

Calibration Procedure For Decade Resistance Model 71-650 Page Navigation
1  2    3    4    5    6  
img
. . . . . . . . . . . .
TB 9-6625-2013-40
SECTION I
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Decade Resistance Standard, Biddle, Model 71-650.  The manufacturer's
manual and purchase specification MIS-10264A were used as the prime data sources in
compiling these instructions. The equipment being calibrated will be referred to as the TI
(test instrument) throughout this bulletin.
a. Model Variations. None.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description.  TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument parameters
Performance specifications
Rero resistance: <0.010
Resistance
Z
Aange: 0.010  to 1.2 M
ccuracy:1  +0.001 ; X0.1 decade and lower
+0.01%; X1.0 decade
+0.0025%; X10 decade and greater
1Obtained
from purchase specification MIS-10264A.
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Reference Calibration
Standards Set, NSN 4931-00-621-7878. Alternate items may be used by the calibrating
activity. The items selected must be verified to perform satisfactorily prior to use and must
bear evidence of current calibration. The equipment must meet or exceed the minimum use
specifications listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio
between the standard and TI.  Where the four-to-one ratio cannot be met, the actual
accuracy of the equipment selected is shown in parenthesis.