Table 7. X10k  Decade Dial Accuracy

Calibration Procedure For Decade Resistance Model 71-650 Page Navigation
  2    3    4    5    6  7
img
. . . . .
TB 9-6625-2013-40
Table 8. X100k  Decade Dial Accuracy
Test instrument
Multimeter indications
Adjustment for
(M )
decade dial
X100K
(fig. 1)
settings
Min
Max
(R)
200000.0100
0.19999501
0.20000501
R2
300000.0100
0.29999251
0.30000751
R3
400000.0100
0.39999001
0.40001001
R4
500000.0100
0.49998751
0.50001251
R5
600000.0100
0.59998501
0.60001501
R6
700000.0100
0.69998251
0.70001751
R7
800000.0100
0.79998001
0.80002001
R8
900000.0100
0.89997751
0.90002251
R9
X00000.0100
0.99997501
1.00002501
R10
1100000.0100
1.09997251
1.10002751
R11
b. Adjustments. Adjust X100K R1 (fig. 1) for a 100.000010 k multimeter indication (R).
15. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective covers on TI.
b. Annotate and affix DA label/form in accordance with TB 750-25.