TB 9-6625-2024-35
Table 1. Calibration Description - Continued
Test Instrument Parameters
Performance Specifications
AND TD-1159/U (Continued)
0.05 s/div to 0.5 s/div
Delayed sweep time
Range:
Accuracy: 3% unmag, 3.5% mag from 50 ms/div to
(over center eight divisions)
0.5 s/div
4% unmag, 4.5% mag from 0.5 s/div to
0.1 s/div and 0.2 s/div to 0.05 s/div
Variable time delay
Range:
0 to 10 times the TIME/DIV OR DLY TIME
control settings from 5 s/div to 1s/div
Differential delay time measurement
Accuracy: 5 s/div to 1 s/div, within 1.4% of
measurement plus 0.3% of FS1
0.5 s/div to 1 s/div, within 0.7%
of measurement plus 0.3% of FS1
1Full
scale is 10 times the TIME/DIV OR DLY TIME setting.
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be
used by the calibrating activity. The items selected must be verified to perform
satisfactorily prior to use and must bear evidence of current calibration. The equipment
must meet or exceed the minimum use specifications listed in table 2. The accuracies
listed in table 2 provide a four-to-one ratio between the standard and TI.
5. Accessories Required. The
accessories required for the
calibration are
common
usage accessories, issued as indicated in paragraph 4 above and are not listed in this
calibration procedure. The following peculiar accessories are also required for this
calibration: Extender Tektronix Type 067-0589-00, and Standardizer, variable from 5 to
80 pF (7916146), and Calibration Adapter, Tektronix Type 067-0587-01.
Table 2. Minimum Specifications of
Equipment Required
Minimum use
Manufacturer and model
Common name
specifications
(part number)
DIGITAL
Range: -3049 to +140 V dc
Hewlett-Packard, Model 3490AOPT060
Accuracy: 0.025%
w/high voltage probe (3490AOPT060)
Must be compatible with required
Tektronix, Type 7B53A or 7B53AN OPT 5
or OPT 11S or TD 1085/U or TD-1159/U
Must be compatible with required
Tektronix, Type 7603 or 7603N OPT 11S
dual time base and vertical
or OS-245 (P)/U
amplifier
See footnote at end of table.