TB 9-6625-2024-35
CALIBRATION PROCESS FOR VERTICAL AMPLIFIER AM-6565/U AND
TEKTRONIX TYPES 7A15A AND 7A15AN OPT 11
a. Remove left side panel from TI and oscilloscope.
b. Install TI into oscilloscope left vertical compartment and dual time base into
horizontal (right) compartment.
c. Position TI controls as listed in (1) through (5) below:
(1) POSITION control to midrange.
(2) POLARITY switch to +UP.
(3) MAG switch to X1.
(4) VOLTS/DIV switch to 10 mV and VARIABLE control fully cw (detent).
(5) AC/GND/DC switch to DC.
d. Press oscilloscope VERT MODE LEFT and TRIG SOURCE LEFT and TRIG
SOURCE VERT MODE pushbuttons.
e. Set dual time base TIME/DIV OR DLY TIME switches to 1 ms and CAL control
fully cw to detent.
f. Energize equipment and allow at least 15 minutes for warm-up.
g. Adjust
POSITION control to
center trace on
oscilloscope center horizontal
graticule line.
h. Rotate VARIABLE control throughout its
range. If
oscilloscope display shifts
more than 0.5 major division, adjust R221 DC BAL (fig. 3). for minimum shift as
VARIABLE control is rotated throughout its range.