TB 9-6625-2043-35
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in paragraph 4 above and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
FREQUENCY
Range: 10 to 11 ms
Fluke, Model PM6681/656
COUNTER
(PM6681/656)
Accuracy: 0.1%
FUNCTION/
Agilent, Model 33250A
ARBITRARY
Amplitude: 0.1 mV to
(33250A)
1.0 V
Range: 1 to 16 V dc
Fluke, Model 8840A/AF05
(AN/GSM-64D)
Accuracy: 0.1%
POWER METER
Range: -10 to 0 dBm
Hewlett-Packard, Model E12-
432A (MIS-30525)
Accuracy: 5%
w/thermistor mount, Hewlett-
Packard, Model H75-478A
(7915907) or 8478B (8478B)
POWER SPLITTER
Range: 1 to 12 GHz
Weinschel, Model 1870A
(7916839)
SYNTHESIZED
Range: 1 to 12 GHz
Anritsu, Model 68369NV
(68369NV)
Accuracy: 1%
Amplitude: +13 dBm
minimum
TRUE RMS
Fluke, Model 8922A/AA
(8922A/AA)
Accuracy: 0.02% (0.5%)
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturers' manuals and TM 11-6625-2857-14 for
this TI.