TB 9-6625-2057-35
Table 3. Frequency Dial Accuracy - Continued
Test instrument
Frequency counter indications
RANGE switch settings
Frequency dial settings
Min
Max
X1K
12
11.64
kHz
12.36
kHz
X10K
1
9.7
kHz
10.3
kHz
X10K
6
38.2
kHz
61.8
kHz
X10K
12
116.4
kHz
123.6
kHz
X100K
1
97
kHz
103
kHz
X100K
6
582
kHz
618
kHz
X100K
12
1.164 MHz
1.236
MHz
b. Adjustments
(1) Set RANGE switch to X1K and frequency dial to 10.
(2) Connect multimeter between TP 4 (fig. 2) and chassis ground.
(3) Adjust R20 (fig. 2) for 0.0 V dc indication on multimeter (R).
(4) Adjust frequency dial to 1 and connect positive lead of multimeter to TP 1 (fig 2).
(5) Adjust R24 (fig. 2) for a -2.0 V dc indication on multimeter (R).
(6) Set frequency dial to 10 and adjust C2 and C8 (fig. 2) for an indication of 10 kHz
(7) Repeat (4) through (6) above to compensate for interaction.
(8) Set RANGE switch to X100K and adjust C4 (fig. 2) for a 1 MHz indication on
frequency counter.
(9) Set RANGE switch to X5 and adjust frequency dial to 1. Repeat technique of
a (2) through (5) above.