TB 9-6625-2059-35
a. Performance Check
(1) Set FREQUENCY MHz switches to 100 000 0 and HI LVL - V X100-NORM
switch to HI LVL.
(2) Adjust BFO-RF LEVEL control until TI 0 dBm light just illuminates.
(3) Connect measuring receiver to TI TRANS-RCVR.
(4) Set measuring receiver to read frequency and then for tuned level measurement.
(5) Measuring receiver will indicate between -2.5 and +2.5 dB in tuned level.
(6) Set HI LVL - VX100-NORM switch to VX100 and adjust BFO-RF LEVEL
control to -100 dBm. If measuring receiver does not indicate between -57.5 and -62.5 dBm,
perform b below.
NOTE
If measuring receiver cannot maintain phase lock, check TI
(7) Repeat technique of (6) above for BFO-RF LEVEL control settings and
FREQUENCY MHz switch settings at
100 000 0 listed in
table 6. Measuring receiver will
indicate within limits specified.
Test instrument
Measuring receiver indications (dB)
BFO-RF LEVEL
FREQUENCY MHz
Min
Max
control settings
switch settings
80
100 000 0
-37.5
-42.5
90
100 000 0
-47.5
-52.5
110
100 000 0
-67.5
-72.5
120
100 000 0
-77.5
-82.5
(8) Set HI LVL - VX100-NORM switch to NORM and BFO-RF LEVEL control
to -80 dBm. Measuring receiver will indicate between 77.5 and 82.5 dB.
(9) Set FREQUENCY MHz switches to 250 000 0 and HI LVL - VX100-NORM
switch to HI LVL.
(10) Adjust BFO-RF LEVEL control until TI 0-dBm light just illuminates.
(11) Repeat (4) above.
(12) Measuring receiver will indicate between -4.0 and +4.0 dB.
(13) Set HI LVL - VX100-NORM switch to VX100 and adjust BFO-RF LEVEL
control to -100 dBm. Measuring receiver will indicate between -56 and -64 dB.
(14) Repeat technique of (13) above for BFO-RF LEVEL control settings and
FREQUENCY MHz switch settings at
250 000 0 listed in
table 7. Measuring receiver will
indicate within limits specified.