SECTION I. IDENTIFICATION AND DESCRIPTION
Table 2. Minimum Specifications of Equipment Required

Calibration Procedure For Rf Millivoltmeter Me-526/Usm (Model 92Bd With Serial Number 2890 And Below) And Models 92B, 92Bd With Serial Number 2891 And Avobe, And 92C Page Navigation
  1    2  3  4    5    6    7    8  
img
. . . . . . . . . .
TB 9-6625-2074-24
Table 1. Calibration Description
Test instrument
parameters
Performance specifications
ME-526/USM
Frequency range: 10 kHz to 1.2 GHz
(Boonton, Model
Accuracy:
r(% of FS + % of reading)
92BD with serial
number 2890 and
FS
+ Reading
below
Voltage range
Frequency range
(%)
(%)
10
to
50
kHz
1
2
100
PV
to  300
mV
10
to
50
kHz
1
1
300
to 3000
mV
50 kHz  to  150
MHz
1
1
100
PV
to 3000
mV
150
to  700
MHz
1
3
100
PV
to 3000
mV
700
to 1200
MHz
1
7
100
PV
to  300
mV
700
to 1200
MHz
1
10
300
to 3000
mV
Boonton, Model
1
1
10 kHz  to  150
MHz
200
PV
to 3000
mV
92BD
3
1
150
to  700
MHz
200
PV
to 3000
mV
(serial number 2891
7
1
700
to 1200
MHz
200
PV
to  300
mV
and above)
10
1
700
to 1200
MHz
300
to 3000
mV
Boonton, Model 92B
10
to
50
kHz
1
2
100
PV
to  300
mV
10
to
50
kHz
1
1
300
to 3000
mV
50 kHz  to  150
MHz
1
1
100
PV
to 3000
mV
150
to  700
MHz
1
3
100
PV
to 3000
mV
700
to 1200
MHz
1
7
100
PV
to  300
mV
700
to 1200
MHz
1
10
300
to 3000
mV
Boonton, Model 92C
10
to
50
kHz
2
2
500
PV
to  300
mV
10
to
50
kHz
2
1
300
to 3000
mV
50 kHz  to  150
MHz
2
1
500
PV
to 3000
mV
150
to  700
MHz
2
3
500
PV
to 3000
mV
700
to 1200
MHz
2
7
500
PV
to  300
mV
700
to 1200
MHz
2
10
300
to 3000
mV
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287 or AN/GSM-705.  Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The accessories listed in table 3 are issued as indicated in
paragraph 4 above and are used in this calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.