TB 9-6625-2094-35
(3) If counter does not indicate between 999 and 1001 Hz, perform b(1) through (4)
below.
(4) Repeat technique of (3) above for FREQUENCY switch settings listed in table 3.
If frequency counter does not indicate within limits specified, perform b(1) through (3)
below.
(5) Position TI controls as listed in (a) through (e) below:
(a) FREQUENCY VERNIER control to 0 kHz.
(b) FREQUENCY switches to indicate 001.000 MHz.
(c) MODULATION MODE switch to FMx1.
(d) MODULATION FREQ switch to DC.
(e) MODULATION FM/AM switch to 10 kHz.
(6) If frequency counter does not indicate between 999.99 and 1020.01 kHz, perform
b(4) through (11) below.
(7) Set MODULATION MODE switch to FMx10.
(8) Frequency counter will indicate between 1089.99 and 1110.01 kHz.
Frequency counter indications
Test instrument
(Hz)
FREQUENCY switch settings
(MHz)
Min
Max
000.001
0999
1001
000.005
4995
5005
000.009
8991
9009
000.010
9990
10,010
000.050
49,950
50,050
000.090
89,910
90,090
000.100
99,999
100,001
000.500
499,995
500,005
000.900
899,991
900,009
001.000
999,990
1,000,010
005.000
4,999,950
5,000,050
009.000
8,999,910
9,000,090
010.000
9,999,900
10,000,100
050.000
49,999,500
50,000,500
090.000
89,999,100
90,000,900
100.000
99,999,000
100,001,000
500.000
499,995,000
500,005,000
(9) Set MODE switch to CW, FREQUENCY switches to 002.000 MHz, and
FREQUENCY VENIER control to +3 kHz. After 1 minute, record frequency counter
indication.
(10) Set FREQUENCY VERNIER control to 0 kHz. After 1 minute, record
frequency counter indication. Subtract from indication recorded in (9) above. Difference
will be between 2500 and 3500 Hz.