TB 9-6625-2114-35
Test instrument
Calibrator indications
RANGE
switch settings
Indications
Min
Max
150
mA
150
145.5 mA
154.5 mA
50
mA
50
48.5 mA
51.5 mA
15
mA
15
14.55 mA
15.45 mA
5
mA
5
4.85 mA
5.15 mA
1.5
mA
1.5
1.455 mA
1.545 mA
.5
mA
.5
.485 mA
.515 mA
150
uA
150
145.5 uA
154.5 uA
50
uA
50
48.5 uA
51.5 uA
15
uA
15
14.55 uA
15.45 uA
5
uA
5
4.85 uA
5.15 uA
1.5
uA
1.5
1.455 uA
1.545 uA
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.