SECTION I. IDENTIFICATION AND DESCRIPTION
Table 1. Calibration Description Continued

Calibration Procedure For Voltmeter, Frequency Selective An/Usm-490(), And Selective Level Meter, Models 3586B And 3586C W/Options 001, 003, And 004 Page Navigation
  1  2  3    4    5    6    7  
img
TB 9-6625-2137-24
Table 1. Calibration Description Continued
Test instrument
parameters
Performance specifications
Amplitude1
Range: 75  /50  input
Accuracy
-80 to +20 dBm
200 Hz to 20 kHz
0.40 dB
20 kHz to 18 MHz
0.20 dB
18 to 32.5 MHz
0.25 dB
-100 to -80 dBm
200 Hz to 20 kHz
0.95 dB
20 kHz to 32.5 MHz
0.75 dB
2
124  input
-80 to +20 dBm
4 to 10 kHz
0.60 dB
10 to 50 kHz
0.50 dB
50 kHz to 5 MHz
0.35 dB
5 to 10 MHz
0.50 dB
2
124  input
-100 dB to -80 dBm
10 to 50 kHz
1.00 dB
50 kHz to 5 MHz
0.75 dB
5 to 10 MHz
1.00 dB
135  , 150  input
2
-80 to +20 dBm
50 kHz to 1 MHz
0.35 dB
10 to 50 kHz
0.50 dB
4 to 10 kHz
0.60 dB
-100 to -80 dBm
50 kHz to 1 MHz
0.75 dB
10 to 50 kHz
1.00 dB
600  input
100 Hz to 108 kHz
-80 to +20 dBm
0.35 dB
-100 to -80 dBm
0.75 dB
Tuned frequency
Range: 200 Hz to 32.5 MHz
Accuracy: 1 X 10-5/year (2 x 10-7)/year w/option 004
Counter
Range: 50 Hz to 32.5 MHz
Accuracy: 1 Hz
Wide band power
Range: 200 Hz to 32.5 MHz
flatness
Accuracy: 200 Hz to 20 kHz
2
dB
20 kHz to 10 MHz
1.0 dB
10 to 32.5 MHz
2  dB
Range: 75  /50  input:
Return loss
50 Hz to 32.5 MHz
30  dB
2
124  input:
4 to 6 kHz
20  dB
6 to 10 kHz
25  dB
10 kHz to 5 MHz
30  dB
135  input:2
4 to 6 kHz
20  dB
6 to 10 kHz
25  dB
10 kHz to 1 MHz
30  dB
600  input:
50 Hz to 108 kHz
25  dB
See footnotes at end of table.