TB 9-6625-2137-35
Table 1. Calibration Description Continued
Test instrument
Performance specifications
parameters
Phase jitter
Demodulated tone frequency range: 960 to 1060 Hz
Accuracy: (10% +.5 p-p)
Input signal level: <30 dB below FS, -65 dBm
Residual range jitter: ≤.5 p-p (50 kHz to 32.5 MHz)
Balance
Range: 50 Hz to 10 MHz
Accuracy: InputFrequency
Balance
124 Ω
10 kHz to 10 MHz -36 dB
135 Ω
10 kHz to 1 MHz
-36 dB
600 Ω
50 Hz to 108 kHz
-40 dB
Adjacent channel
Accuracy: 75 dB minimum 2850 Hz
rejection
Carrier channel
Accuracy: 30 dB 45 Hz
rejection
60 dB 90 Hz
Accuracy: 60 dB 1100 Hz
Range: Bandwidth 3100 Hz
Accuracy: 60 dB 1850 Hz
Range: Bandwidth 400 Hz 1100 Hz at 60 dBm 0 rejection
Pilot (20 Hz) filter bandwidths:
Range: 45 Hz at 30 dBm 0 rejection
Range: 90 Hz at 60 dBm 0 rejection
Band pass flatness
Accuracy: 0.3 dB
Range: Bandwidth 400 Hz 50 Hz
Accuracy: 0.3 dB
Range: Bandwidth 20 Hz 3 Hz
Accuracy: 0.3 dB
1Range
specifications calibrated to -100 dBm to +20.
2Not applicable for model 3586C.
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287, or AN/GSM-705. Alternate items may be used by the
calibrating activity. The items selected must be verified to perform satisfactorily prior to
use and must bear evidence of current calibration. The equipment must meet or exceed the
minimum use specifications listed in table 2. The accuracies listed in table 2 provide a four-
to-one ratio between the standard and TI. Where the four-to-one ratio cannot be met, the
actual accuracy of the equipment selected is shown in parenthesis.
5. Accessories Required. The
accessories listed in
table 3 are issued as
indicated in
paragraph 4 above and
are used in this
calibration procedure. When necessary, these items
may be substituted by equivalent items, unless specifically prohibited.