TB 9-6625-2137-35
Figure 3. Balance test 600 Ω -
equipment setup.
(5) Press TI MEASUREMENT/ENTRY
[OFFSET]
and
ENTRY
[RDNG
OFFSET] pushbuttons.
(6) Disconnect cable No. 1 from TI 75 Ω input.
(7) Press TERMINATION [600 Ω] pushbutton. TI will indicate less than or equal
to -40 dBm O.
(8) Reconnect cable No. 1 and repeat (2) through (7) above at 100 kHz.
(9) Remove balance test 600 Ω adapter from setup.
NOTE
Steps (10) through (20) do not apply for model 3586C.
NOTE
Cable No. 1 must be the same length and capacitance for 10
MHz balance check.
10