TB 9-6625-2138-35
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
Table 2. Minimum Specifications of
Equipment Required
Common name
Minimum use specifications
Manufacturer and model (part number)
Range: 105 to 125 V ac
General Radio, Model W10MT3AS3 or
Ridge, Model 9020F (7910809)
Accuracy: 1%
DIFFERENTIAL
Range: -23 to +15.5 V dc
John Fluke, Model 887AB/AN (887AB/AN)
Accuracy: 0.08%
Range: 74 to 9603 Hz
Hewlett-Packard, Model AN/USM459B
(AN/USM459B)
Accuracy: 0.007%
Range: -12 to +12 V peak
Tektronix, Type 2465BOPT46
(2465BOPT46)
Accuracy: 3%
Range: 0 to 1 kHz, 5 V peak
LeCroy, Model 9210 (9210) w/plug-ins,
Duty cycle: Adjustable to 40%
LeCroy, Models 9211 (9211) and 9215
(9215)
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration
process. Personnel should become familiar with the entire bulletin before beginning
the calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.
d. When indications specified in
paragraphs 8 through 12 are not within
tolerance,
perform the power supply check prior to making adjustments. After adjustments are
made, repeat paragraphs 8 through 12. Do not perform power supply check if all
other parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.