TB 9-6625-2139-35
(53) Repeat technique of step (52) above for A TRIGGER pushbutton combinations
Table 23. A
Trigger Level A
Source to EXT INPUT
Test instrument
Test instrument
Test instrument
A TRIGGER
A TRIGGER LEVEL with
adjustments
pushbutton combinations
TRIG VIEW
in stable display test
MODE
SLOPE
YES
NO
b
NORM
IN:
b
P-P AUTO
IN:
b
P-P AUTO
OUT:
b
(54) Release TRIG VIEW pushbutton.
(55) Pull X10 CAL control to out position and press A TRIGGER NORM
pushbutton.
(56) Set oscilloscope calibrator LEVEL SINE output to 120 mV and 60 MHz.
(57) Press in and hold TRIG VIEW pushbutton while adjusting A TRIGGER
LEVEL control to obtain a stable display.
(58) Repeat technique of step (57) above for A TRIGGER pushbutton combinations
Table 24. A
Trigger Level A
Source to EXT INPUT
Test instrument
Test instrument
A TRIGGER LEVEL with
A TRIGGER
TRIG VIEW
Test instrument
pushbutton combinations
in stable display test
adjustments
MODE
SLOPE
Yes
No
b
NORM
IN:
b
P-P AUTO
IN:
b
P-P AUTO
OUT:
b
(59) Release TRIG VIEW pushbutton.
(60) Set oscilloscope calibrator LEVEL SINE output to 150 mV (200 mV for type
2235) and 100 MHz output.
(61) Press A TRIGGER NORM pushbutton. Next press in and hold TRIG VIEW
pushbutton and adjust A TRIGGER LEVEL control to obtain a stable display.
(62) Repeat technique of step (61) above for A TRIGGER pushbutton combinations
Table 25. A
Trigger Level A
Source to
EXT INPUT
Test instrument
Test instrument
A TRIGGER LEVEL with
A TRIGGER
TRIG VIEW
Test instrument
pushbutton combinations
in stable display test
adjustments
MODE
SLOPE
YES
NO
B
NORM
IN:
B
P-P AUTO
IN:
B
P-P AUTO
OUT:
B
30