TB 9-6625-2144-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Radar Test Set, AN/UPM-29A, TS-147B/UP, and TS-1335A.
The
manufacturers' manuals, TM 11-6625-917-15 and TM 11-1247 (Series) were used as the
prime data sources in compiling these instructions. The equipment being calibrated will be
referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Differences among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 3
hours using the microwave technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications
parameters
TS-147 ( )/UP
AN/UPM-29 ( ) and TS-1335A
Range: 8470 to 9630 MHz
Range:
15.7 to 17.0 GHz
8 MHz absolute
Accuracy: 2.5 MHz absolute
Accuracy:
1MHz relative for
1.5 MHz/60 MHz
increments of less than 60
increment relative
MHz 1 MHz at 9310
Range: 8500 to 9600 MHz
Range: 15.7 to 17.0 GHz
RF output power
Range: -7 to -85 dBm
Range: -43 to -90 dBm
Accuracy: 1.5 dB
Accuracy: 2 dB
RF input power1
Range: +7 to +30 dBm
Range: +3 to +30 dBm
Accuracy: 1.5 dB
Accuracy: 1.5 dB
1Item
should be calibrated to highest cardinal point allowed by standard (Example: +7, +10, +20, +30).