• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 4. Dc Voltage - Continued
Table 5. Dc Current - Continued

Calibration Procedure For Ac And Dc Voltmeters And Ac And Dc Ammeters (General) Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. .
TB 9-6625-2146-24
Figure 1. Dc current - equipment setup.
Table 5. Dc Current
Multimeter indications (mV dc)
Simpson, Model 13291
SRI, Models C-CILKI and
ME-221/U and Weston,
Test
C-CILORD
Models 430, 901, 931, and 622
instrument
series
range
0.25%
0.5%
2%
Min
Max
Min
Max
Min
Max
---
---
49.75
50.25
49
51
5
PA
---
---
99.5
100.5
98
102
10
PA
---
---
14.925
15.075
14.7
15.3
15
PA
---
---
19.90
20.10
19.6
20.4
20
PA
---
---
29.85
30.15
29.4
30.6
30
PA
---
---
49.75
50.25
49
51
50
PA
---
---
99.5
100.5
98
102
100
PA
---
---
14.925
15.075
14.7
15.3
150
PA
---
---
19.90
20.10
19.6
20.4
200
PA
---
---
49.75
50.25
49
51
500
PA
---
---
99.5
100.5
98
102
1000
PA
0.03
mA
---
---
29.85
30.15
29.4
30.6
0.05
mA
49.875
50.125
---
---
49
51
0.1
mA
99.75
100.25
99.5
100.5
98
102
0.2
mA
---
---
19.90
20.10
19.6
20.4
0.3
mA
---
---
29.85
30.15
29.4
30.6
0.5
mA
49.875
50.125
49.75
50.25
49
51
See footnote at end of table
8 CHANGE 1


Privacy Statement
Press Release
Contact

© Copyright Integrated Publishing, Inc.. All Rights Reserved.