TB 9-6625-2146-35
SECTION I
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Ac and Dc Voltmeters and Ac and Dc Ammeters listed in table 1. The
manufacturers manuals and/or TM's were used as the prime data sources in compiling
these instructions. The equipment being calibrated will be referred to as the TI (test
instrument) throughout this bulletin.
a. Model Variations. Variations among models are listed in the text.
b. Time and Technique. The time required for each calibration is approximately 1
2. DA Form 2416 (Calibration Data Card). Forms, records, and reports required for
calibration personnel at all levels are prescribed by TB 750-25-1. DA Form 2416 must be
annotated in accordance with TB 750-25-1 for each calibration performed.
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument
Manufacturer and model
Parameters
Performance specifications
General Electric, Model AK-4
Ac volts
Range: 0 to 750 V
3% FS at 60 Hz
Accuracy:
Ac amps
Range: 0 to 800 A
3% FS at 60 Hz
Accuracy:
Sensitive Research Instrument
Dc milliamps
Range: 0 to 100 mA
Corporation (SRI), Model C-CILKI
0.25% of FS
Accuracy:
(horizontal position)
Sensitive Research Instrument
Dc amps
Range: 0 to 30 A
Corporation (SRI), Model C-CILORD
0.25% of FS
Accuracy:
(horizontal position)
ME-221/U
Dc milliamps
Range: 0 to 500 mA
(horizontal position)
0.5% of FS
Accuracy:
ME-489/U
Ac volts
Range: 0 to 150, 300, 600 V
3% of FS at 50 or 60 Hz
Accuracy:
Ac amps
Range: 0 to 6, 15, 60, 150, 300 A
3% of FS at 50 or 60 Hz
Accuracy:
ME-65/U and ME-65A/U
Ac amps
Range: 0 to 200 A
(horizontal position)
0.5% of FS, 20 to 500 Hz
Accuracy:
Westinghouse, Model PA-5
Ac amps
Range: 0 to 200 A
(horizontal position)
0.5% of FS, 20 to 135 Hz
Accuracy:
Westinghouse, Model PY-4
Ac volts
Range: 0 to 450 V
Style 936-368 (horizontal position)
0.75% of FS, 5 to 200 Hz
Accuracy:
TS-340/U (horizontal position)
Ac volts
Range: 0 to 750 V
0.5% of FS, 25 to 400 Hz
Accuracy;
Dc volts
Range: 0 to 750 V
0.5% of FS
Accuracy:
2