Figure 1. Receiver board A3 (A13) - adjustment locations.
Filters

Calibration Procedure For Transmission Impairment Test Set An/Usm-485 And Model 4935A With Options 001, 002, And 003 Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . . . . . .
TB 9-6625-2151-24
9. Transmitter Flatness
a. Performance Check
f
(1) Connect TI left TRMT/RCV 310 to multimeter Input HI and LO using 600
eedthrough termination.
(2) Press pushbuttons and keys as listed in (a) through (d) below:
DISPLAY key to select TRMT.
(a)
NOR pushbutton.
(b)
SET UP TRMT 600  pushbutton.
(c)
IMPULSE NOISE DISPLAY F2 1004 key.
(d)
(3) Adjust OUTPUT LEVEL control for a 2.45 V ac indication on multimeter.
(4) Press  , STEP UP, STEP DOWN, and
keys for TI right display indications
listed in table 4. Multimeter indications will be within limits specified.
Table 4. Transmitter Flatness at +10 dBm
Test instrument
Multimeter indications
right display
(Vac)
indications
Min
Max
20  Hz
2.183
2.748
10  kHz
2.394
2.507
60  kHz
2.312
2.595
85  kHz
2.312
2.595
(5) Press IMPULSE NOISE DISPLAY F2 1004 key.
(6) Adjust OUTPUT LEVEL control for a 0.775 V ac indication on multimeter.
(7) Press  , STEP UP, STEP DOWN, and
keys for TI right display indications
listed in table 5. Multimeter indications will be within limits specified.
Table 5. Transmitter Flatness at 0 dBm
Test instrument
Multimeter indications
right display
(Vac)
indications
Min
Max
20  Hz
0.690
0.869
10  kHz
0.757
0.793
60  kHz
0.731
0.820
85  kHz
0.731
0.820
(8) Press IMPULSE NOISE DISPLAY F2 1004 key.
(9) Adjust OUTPUT LEVEL control for a 7.75 mV ac indication on multimeter.
(10) Press  , STEP UP, STEP DOWN, and
keys for TI right display indications
listed in table 6. Multimeter indications will be within limits specified.
8