TB 9-6625-2153-35
Table 9. Cornell-Dubilier, Model RDA
Test instrument
Performance specifications
Range: 1 Ω to 10 Ω
Accuracy: 5%
Range: 10 Ω to 100 Ω
Accuracy: 3%
Decade
Decade
switch
switch positions
Min
Max
0 to 10
1
00.95000
01.05000
2
01.90000
02.10000
3
02.85000
03.15000
4
03.80000
04.20000
5
04.75000
05.25000
6
05. 70000
06.30000
7
06.65000
07.35000
8
07.60000
08.40000
9
08.55000
09.45000
10
09.50000
10.50000
0
---
---
0 to 100
1
09.70000
10.30000
2
19.40000
20.60000
3
29.10000
30.90000
4
38.80000
41.20000
5
48.50000
51.50000
6
58.20000
61.80000
7
67.90000
72.10000
8
77.60000
82.40000
9
87.30000
92.70000
10
97.00000
103.00000
0
---
---
Table 10. ESI, Models DB-62 and DB-877
Test instrument
Performance specifications
ESI, Models DB-62 and
DB 877
Range1
Accuracy
DB-62
DB-877
0.01Ω/step
5%
10%
0.1Ω/step
0.5%
1%
1Ω/step
0.07%
0.12%
10Ω/step
0.025%
0.03%
100Ω/step
0.02%
0.02%
1000Ω/step
0.02%
0.02%
10,000Ω/step
0.02%
0.02%
100,000Ω/step
0.02%
0.02%
1,000,000Ω/step
0.02%
0.02%
See footnote at end of table.
17