TB 9-6625-2153-35
Table 10. ESI, Models DB-62 and DB-877 - Continued
Test instrument
Megohms
Decade
Decade switch
DB-62
DB-877
switch
positions
Min
Max
1
0.999800
1.000200
1 MΩ/step
2
1.999600
2.000400
3
2.999400
3.000600
4
3.999200
4.000800
5
4.999000
5.001000
6
5.998800
6.001200
7
6.998600
7.001400
8
7.998400
8.001600
9
8.998200
9.001800
10
9.998000
10.002000
0
---
---
1Use
appropriate ranges for particular model of TI being tested.
Table 11. General Radio, Model 1433 Series
Test instrument
Performance Specifications
1433 Series
Range1
Accuracy
0.01 Ω/step
2%
0.1 Ω/step
0.4%
1 Ω/step
0.1%
10 Ω/step
0.04%
100 Ω and above/step
0.01%
Decade switch
Decade switch
Min
Max
positions
Ohms
1
00.00980
00.01020
0.01Ω/step
2
00.01960
00.02040
3
00.02940
00.03060
4
00.03920
00.04080
5
00.04900
00.05100
6
00.05880
00.06120
7
00.06860
00.07140
8
00.07840
00.08160
9
00.08820
00.09180
10
00.09800
00.10200
0
- - -
- - -
1
00.09960
00.10040
0.1Ω/step
2
00.19920
00.20080
3
00.29880
00.30120
4
00.39840
00.40160
5
00.49800
00.50200
6
00.59760
00.60240
See footnote at end of table.
20