TB 9-6625-2153-35
Table 16. Leeds and Northrup, Models 4755 and 4776 - Continued
Test instrument
4755
4776
Decade
Decade switch
switch
Min
Max
Min
Max
position
100Ω/step(cont)
7
699.64500
700.35500
699.29000
700.71000
8
799.59500
800.40500
799.19000
800.81000
9
899.54500
900.45500
899.09000
900.91000
10
999.49500
1000.50500
---
---
0
---
---
---
---
Kilohms
1 kΩ/step
1
0.9994950
1.0005050
0.9989900
1.0010100
2
1.9989950
2.0010050
1.9979900
2.0020100
`
3
2.9984950
3.0015050
2.9969900
3.0030100
4
3.9979950
4.0020050
3.9959900
4.0040100
5
4.9974950
5.0025050
4.9949900
5.0050100
6
5.9969950
6.0030050
5.9939900
6.0060100
7
6.9964950
7.0035050
6.9929900
7.0070100
8
7.9959950
8.0040050
7.9919900
8.0080100
9
8.9954950
9.0045050
8.9909900
9.0090100
10
9.9949950
10.005005
---
- - -
0
---
---
---
- - -
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.