TB 9-6625-2164-35
a. Performance Check
(1) Press MENU/TEST and MEMORY DISPLAY 4, 5, 6 and MENU/TEST
pushbuttons.
(2) Set CH 1 VOLTS/DIV switch to 100 mV.
(3) Set TIME/DIV switch to 10 s.
(4) Connect function/arbitrary generator Output to CH 1 input, using 50 Ω
termination, and press function/arbitrary generator Sine key.
(5) Set function/arbitrary generator frequency for 50 kHz and amplitude for 8
divisions on TI display.
(6) Set leveled sine frequency to 70 MHz.
(7) Set TIME/DIV to 50 ns. TI display will indicate at least 5.7 divisions.
(8) Move connection from TI CH 1 input to CH 2 input.
(9) Press AQR MODE CH 2→2 pushbutton and repeat (2) through (7) above for CH 2.
(10) Set TIME/DIV switch to 500 s.
divisions centered on TI display.
(12) Set TIME/DIV switch to 50 ms.
(13) Set function/arbitrary generator frequency to 10 Hz. TI will indicate at least 5.7 divisions.
(14) Move connection from TI CH 2 input to CH 1 input.
(15) Press AQR MODE CH 1→1 pushbutton and repeat (10) through (13) above.
(16) Remove all connections from TI.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Press MENU/TEST and MEMORY DISPLAY 4, 5, 6 and MENU/TEST pushbuttons.