SECTION III. CALIBRATION PROCESS FOR RADIO TEST SET, TS3951/PRM34
Frequency

Calibration Procedure For Radio Test Set, Ts3951/Prm34, Receiver Test Set, An/Arm-186 And Collins, Model 972Q-4 Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . . . . .
TB 9-6625-2171-24
b.  Adjustments. Adjust R41 (fig. 1) for overall output levels and C29 (fig. 1) for
higher frequency output levels for spectrum analyzer indications between -94 and -100 dBm
for SN 1 through 364 and between -95 and -101 dBm for SN 635 and above (R).
Figure 1. Test instrument - bottom view.
9. Power Measurement
a. Performance Check
(1) Connect equipment as shown in figure 2.
Figure 2. Power and frequency check equipment setup.
6