TB 9-6625-2173-24
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description.  TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Table 1. Calibration Description
Test instrument parameters
Performance specifications
Tektronix, Type SC501
Vertical bandwidth
dc to > 5 MHz
Vertical sensitivity
Range: 10,100 and 1000 mV/div
Accuracy:
r3%
Horizontal sweep
Range: 1 Ps to 100 ms in a 10 to 1 ratio
Accuracy: r5%
Magnifier: X5 r5%
Tektronix, Type SC502
Vertical bandwidth
dc to >15 MHz
Vertical sensitivity
Range: 1 mV to 20 V/div; 14 steps in a 1-2-5-sequence
Accuracy:
5 mV to 20 V/div: r2%
1 and 2 mV/div: r5%
Horizontal sweep
Range: 0.2 Ps to .5 s/div: 20 steps in a 1-2-5 sequence
Accuracy:
Measured over center 8 divisions
50 ms/div to 1 ms/div: r2% unmagnified; r3% magnified
0.5 s/div to 0.1 s/div, and 0.5 Ps/div to 0.2 Ps/div:
r3% unmagnified; r4% magnified
Calibrator voltage
Range: 0.6 V square wave
Accuracy:
r1%
Tektronix, Type SC504
Vertical bandwidth
dc to > 80 MHz
Vertical sensitivity
Range: 5 mV to 10 V/div; 11 steps in a 1-2-5-sequence
Accuracy:
r2%
Aberrations
r5%
3