Table 3. Voltage Accuracy
Figure 2. View of underside of module assembly.

Calibration Procedure For True Rms Voltmeter Me-545/G (Model 5002-S-8040) Page Navigation
  1    2    3    4  5  6    7    8  
img
. . . .
TB 9-6625-2193-24
Table 3. Voltage Accuracy - Continued
Test instrument
Calibrator output
Test instrument indications
Range
Filter
Voltage
Frequency
Min
Max
Adj. (R)
31.62  V
On
30
V
20
Hz
28.50  V
31.50  V
---
31.62
V
30
V
1.0
kHz
29.10
V
30.90
V
---
31.62
V
Off
30
V
50
kHz
29.10
V
30.90
V
---
31.62
V
30
V
500
kHz
28.50
V
31.50
V
---
31.62
V
20
V
1.0
MHz
18.00
V
22.00
V
---
100.0
V
On
100
V
20
Hz
95.0
V
105.0
V
---
100.0
V
100
V
1.0
kHz
97.0
V
103.0
V
---
100.0
V
Off
100
V
50
kHz
97.0
V
103.0
V
---
100.0
V
40
V
500
kHz
38.0
V
42.0
V
---
RESET calibrator and connect power amplifier to setup.
316.2
V
On
300
V
40
Hz
285.0
V
315.0
V
---
316.2
V
300
V
1.0
kHz
291.0
V
309.0
V
---
316.2
V
Off
300
V
50
kHz
291.0
V
309.0
V
---
316.2
V
300
V
100
kHz
285.0
V
315.0
V
---
RESET calibrator and disconnect power amplifier from setup.
1Record
TI indications.
2Figure
1.
3Figure 2.
4C5 and C6 interact. Repeat if C6 is adjusted.
Figure 1. Top view of module assembly.
CHANGE 1
6