SECTION I. IDENTIFICATION AND DESCRIPTION
Table 1. Calibration Description - Continued

Calibration Procedure For Function Generator Sg-1288/G Page Navigation
  1  2  3    4    5    6    7  
img
TB 9-6625-2233-24
Table 1. Calibration Description - Continued
Test instrument
Performance specifications
parameters
Sine wave3
Output amplitude: <10 V p-p
Total harmonic
Range: 2 mHz to 19.99 Hz
distortion
Accuracy:  <1% (-40 dBc)
Range: 20 Hz to 99.9 kHz
Accuracy:  <0.5% (-46 dBc)
Range: 100 to 999.9 kHz
Accuracy:  <1 % (-40 dBc)
Range: 1 to 5.999 MHz
Accuracy:  <2% (-34 dBc)
Range: 6 to 7.999 MHz
Accuracy:  <2.5% (-32 dBc)
Range: 8 to 9.999 MHz
Accuracy:  <3% (-30 dBc)
Range: 10 to 19.99 MHz
Accuracy:  <4% (-28 dBc)
Output amplitude: >l0 V p-p
Range: 2 mHz to 19.99 Hz
Accuracy:  <1 % (-40 dBc)
Range: 20 Hz to 99.9 kHz
Accuracy:  <0.5% (-46 dBc)
Range: 100 to 999.9 kHz
Accuracy:  <1 % (-40 dBc)
Range: 1 to 5.999 MHz
Accuracy:  <2% (-34 dBc)
Range: 6 to 7.999 MHz
Accuracy:  <3% (-30 dBc)
Range: 8 to 9.999 MHz
Accuracy:  <4.5% (-27 dBc)
Range: 10 to 19.99 MHz
Accuracy:  <5% (-26 dBc)
Sine wave4
Range: <1 MHz
Accuracy:  <-60 dBc
Non-harmonically
related signals 200 Hz
Range: 1 to 9.99 MHz
of output
Accuracy:  <-40 dBc
Range: 10 to 19.99 MHz
Accuracy:  <-36 dBc
Square wave
Rise/falltime: <13 ns
Aberrations: <5%, 20 mV of p-p amplitude
Symmetry: 10 to 90% in l% steps to l MHz linearly decreasing to 50%
fixed 20 MHz
Accuracy:  (l% +20 ns)
See footnotes at end of table.