Table 1. Calibration Description - Continued
Table 1. Calibration Description - Continued

Calibration Procedure For Function Generator Sg-1288/G Page Navigation
  1    2  3  4    5    6    7    8  
img
TB 9-6625-2233-24
Table 1. Calibration Description - Continued
Test instrument
Performance specifications
parameters
Triangle wave
Linearity: (10 to 90%):5
Range: 2 mHz to 99.99 kHz
Accuracy:  1%
Range: 100 kHz to 1.99 MHz
Accuracy:  2%
Range: 2 to 5 MHz
Accuracy:   10%
Symmetry: 10 to 90% in 1% steps to 1 MHz linearly decreasing to 5 MHz
Accuracy:  (1 % +20 ns)
Frequency modulation
Range: 10 MHz carrier, 1 kHz modulation frequency at 100 kHz
distortion
deviation
Accuracy:  <-35 dBc
Sine wave
Range: 2 mHz to 99.99 kHz
amplitude6
Range: 1 to 30 V7
Accuracy:  (2% +10 m V)
Range: 100 mV to 1 V
Accuracy:  (2% +2 mV)
Range: 1 to 100 mV
Accuracy:  (2% +1 mV)
Range: 100 kHz to 999.9 kHz
Range: 1 to 30 V
Accuracy:  (4% +10 mV)
Range: 100 mV to 1 V
Accuracy:  (4% +2 mV)
Range: 1 to 100 mV
Accuracy:  (4% +1 mV)
Range: 1 to 4.999 MHz
Range: 1 to 30 V
Accuracy:  (5% +10 mV)
Range: 100 mV to 1 V
Accuracy:  (5% +2 mV)
Range: 1 to 100 mV
Accuracy:  (5% +1 mV)
Range: 5 to 15.99 MHz
Range: 1 to 30 V
Accuracy:  (12%+10 mV)
Range: 100 mV to 1 V
Accuracy:  (12%+2 mV)
Range: 1 to 100 mV
Accuracy:  (12% +1 mV)
Range: 16 to 19.99 MHz
Range: 1 to 30 V
Accuracy:  (30% +10 mV)
See footnotes at end of table.
4