TB 9-6625-2233-35
Table 13. Square Wave Amplitude 100 kHz to 4.999 MHz- Continued
Measuring receiver indications
Output
(mV)
(Hz)
(V p-p)
Min
Max
1
M
.099
45.03
53.97
1
M
.1
45.00
55.00
4.999 M
.099
45.03
53.97
4.999 M
.1
45.00
55.00
(12) Reduce TI output to minimum. Disconnect UNBALANCED output from
measuring receiver power sensor and connect UNBALANCED output to oscilloscope
Vertical 1 input using a 50 Ω feedthrough termination.
(13) Press pushbuttons as listed in (a) through (h) below:
(a)
FREQUENCY.
(b)
1.
(c)
EXP.
(d)
6.
(e)
ENTER.
(f)
AMPLITUDE.
(g)
.999.
(h)
ENTER.
(14) Set oscilloscope Vertical 1 inputs for DC Coupling, 1MΩ Input, scaling for
approximately 4 or 5 divisions displayed, sweep speed for approximately 4 or 5 cycles
displayed and volts pk-pk measurement.
(15) Oscilloscope will indicate within minimum maximum limits as listed in table 14.
(16) Repeat technique of (13) through (15) above substituting settings and
Table 14. Square Wave Amplitude 1 to
19.99 MHz
Oscilloscope indications
Output
(V p-p)
(Hz)
(V p-p)
Min
Max
1
M
.999
.935 mV
1.06
V
1
M
1
.930 mV
1.07
V
1
M
9.99
9.38
V
10.6 V
1
M
10
9.38
V
10.6
V
1
M
15
14.1
V
15.9
V
4.999 M
.999
.935 mV
1.06
V
4.999 M
1
.930 mV
1.07
V
4.999 M
9.99
9.38
V
10.6 V
4.999 M
10
9.38
V
10.6
V
4.999 M
15
14.1
V
15.9
V
5
M
.009
5.92
mV
12.1
mV
5
M
.01
5.8
mV
14.2
mV
5
M
.099
84
mV
114
mV
5
M
.1
84
mV
116
mV
5
M
.999
875
mV
1.12
V
5
M
1
870
mV
1.13
V
5
M
9.99
8.78
V
11.2
V
23