TB 9-6625-2239-35
Test instrument
parameters
Performance specifications
Vertical deflection
Range: 5 mV to 5 V per division
Accuracy: 3% (2% for AN/USM-425(V)l)
Timing
Range:
Unmag: 0.5 s/div to 0.05 s/div (0.5 s/div to 0.02 s/div for Tektronix, Type
A sweep
465B)
Mag: X10 extends range to 5 ns/div (2 ns/div for Tektronix, Type 465B)
Accuracy:
Unmag: 2%
Mag: 3%
B sweep
Range:
Unmag: 50 ms/div to 0.05 s/div (50 ms/div to 0.02 s/div for Tektronix, Type
465B)
Mag: Xl0 extends range to 5 ns/div (2 ns/div for Tektronix, Type 465B)
Accuracy:
Unmag: 2%
Mag: 3%
Range: 5 s to 0.2 s after start of A sweep
Delay time and
differential time
Accuracy: 1% over one or more major dial divisions
measurement
0.01 major dial division over less than 1 major dial division
Range: 3.5 ns or less
Risetime
Aberrations: +3%, -3%, total p-p 3% (+4%, -4%, total p-p 4% for Tektronix, Type
465B on 5 mV to 2 V range)
Range: Dc to 100 MHz
Calibrator
Range: 0.3 V (1.0 V for AN/USM-425(V)1)
Accuracy: 1.0%
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2
identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287, or AN/GSM-705. Alternate items may be used
by the calibrating activity when the equipment listed in table 2 is not available. The items
selected must be verified to perform satisfactorily prior to use and must bear evidence of
current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI. Where the four-to-one ratio cannot be met, the four-to-one accuracy will
be listed, and the actual accuracy of the equipment selected will be shown in parenthesis.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
5 - 80 pF standardizer.