TB 9-6625-2239-35
Tektronix, Type 465B) perform appropriate adjustment for model identified in table 10.
Rise time will be 3.5 ns or less.
NOTE
Make adjustments in order listed in table 10 below for
optimum flat-top square wave.
(11) Press VERT MODE CH 1 pushbutton to on (in).
(12) Repeat technique of (8) through (10) above for CH 1. If square wave displayed
exceeds tolerances listed in (10) above, perform appropriate adjustments for model
identified in table 10. Risetime will be 3.5 ns or less.
CH 1
CH 2
Model
adjustments (R)
adjustments (R)
Figure
AN/USM-425(V)l
C4173,
R4173,
C4174, R4406, C4405, R4410,
1
R4174, R4167, R4106
C4410, C4342, C4412,
C4447, R4447
Tektronix, Type 465
C27, C33, C105, R105, C77, C83, C205, C249,
2
C122, R122, C149, R149
R249
---
C442, R442, C443 (C4422,
4
R4438, C4438 for discrete
version)
Tektronix, Type 465B
R1190, C1192, R1335,
R1685, Cl690, R1634,
3
C1342
C1643, C2315 (C2308
below
SN
B030000),
R2325 (R2312 below SN
B030000), and R2243
(R2245
below
SN
B030000), and C2346
(C2350
below
SN
B030000)
(13) Remove connection from oscilloscope calibrator and TI.
(14) Connect oscilloscope calibrator SOURCE/MEASURE CHAN 1 to CH 1 through
a 50 Ω feedthrough termination.
(15) Press VERT MODE CH 1 pushbutton to on (in) and set VERT VOLTS/DIV to 50mV.
(16) Set A AND B TIME/DIV switches to .1 ms, and press and release Xl0 MAG
pushbutton to off (out).
(17) Set oscilloscope calibrator LEVEL SINE output to 50 kHz, and amplitude for 6
divisions of display on TI crt.
NOTE
To perform step below; press oscilloscope calibrator EDIT
FIELD pushbutton as required to place underline under one of
(18) Rotate oscilloscope calibrator knob
below EDIT FIELD pushbutton to
sweep
17