Table 6. Model 75 Series II Specifications
Table 7a. Model 75 Series III Dc Voltage

Calibration Procedure For Multimeter, Models 73, 73 Series Iii, 75, 75 Series Ii, 75 Series Iii, 77, 77 Series Ii, 77 Series Iii, 77/An, And 77/Bn Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
TB 9-6625-2243-24
Table 6b. Model 75 Series II Ac Voltage
Performance Checks
Calibrator output
Test
settings
Test instrument indications
instrument
Voltage
Frequency
range settings
(V)
(Hz)
Min
Max
3.2 V
3
45
2.938
3.062
3.2 V
3
500
2.938
3.062
32
V
30
45
29.38
30.62
32
V
30
1000
29.38
30.62
320
V
300
1000
293.8
306.2
320
V
300
45
293.8
306.2
750
V
700
45
684
716
750
V
700
1000
684
716
Table 6c. Model 75 Series II Resistance
Performance Checks
Calibrator
Test instrument
Error
Output
range settings
display indications
settings
( )
(%)
( )
1
320
100
0.7
3200
1k
0.6
32  k
10  k
0.6
2
320  k
100  k
0.6
3.2 M
1M
0.6
32  M
10  M
2.1
1
Set calibrator 2 wire Comp ON.
2
Set calibrator 2 wire Comp OFF.
Table 6d. Model 75 Series II Dc Current
Performance Checks
Test instrument
Calibrator
Test instrument indications
range settings
output settings
Min
Max
32  mA
30 mA
29.53
30.47
320  mA
300 mA
295.3
304.7
1
10  A
5A
4.9
5.1
1
Connect calibrator amplifier CURRENT OUTPUT HI and LO to TI 10A and COM.