Table 7a. Model 75 Series III Dc Voltage
Table 8b. Model 77 Ac Voltage

Calibration Procedure For Multimeter, Models 73, 73 Series Iii, 75, 75 Series Ii, 75 Series Iii, 77, 77 Series Ii, 77 Series Iii, 77/An, And 77/Bn Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
.
TB 9-6625-2243-24
Table 7c. Model 75 Series III Resistance
Performance Checks
Calibrator
Test instrument
Error
Output
range settings
display indications
settings
( )
(%)
( )
1
320
100
0.7
3200
1k
0.6
32  k
10 k
0.6
2
320  k
100 k
0.6
3.2 M
1M
0.6
32  M
10 M
2.1
1
Set calibrator 2 wire Comp ON.
2
Set calibrator 2 wire Comp OFF.
Table 7d. Model 75 Series III Dc Current
Performance Checks
Test instrument
Calibrator
Test instrument indications
range settings
output settings
Min
Max
32 mA
30 mA
29.53
30.47
320 mA
300  mA
295.3
304.7
1
10 A
5A
4.9
5.1
1
Connect calibrator amplifier CURRENT OUTPUT HI and LO to TI 10A and COM.
Table 8. Model 77 Specifications
Performance specifications
Test instrument parameters
(% of reading + counts)
Dc voltage
Range:
Accuracy:
(0.3% + 1)
320 mV to 320 V
(0.4% + 1)
1000 V
Ac voltage
Range:
Accuracy:
(2.0% + 2)
3.2 V, 45 to 500 Hz
(2.0% + 2)
32 to 750 V, 45 Hz to 1 kHz
Dc Current
Range:
Accuracy:
(1.5% + 2)
32 mA
(2.0% + 2)
320 mA
(1.5% + 2)
10 A
1
Ac Current
Range: 0 to 10 A, 45 Hz to 1 kHz
Accuracy:  (3.0% + 2)
Resistance
Range:
Accuracy:
(0.5% + 2)
320
(0.5% + 1)
3200  to 3.2 M
(2.0% + 1)
32 M
1
Ac current verified by dc current checks because both use the same shunt resistors.
12