Table 10. Model 77 Series III Specifications
Table 11a. Model 77/AN Dc Voltage

Calibration Procedure For Multimeter, Models 73, 73 Series Iii, 75, 75 Series Ii, 75 Series Iii, 77, 77 Series Ii, 77 Series Iii, 77/An, And 77/Bn Page Navigation
  10    11    12    13    14  15  16    17    18  
img
TB 9-6625-2243-24
Table 10a. Model 77 Series III Dc Voltage
Performance Checks
Test
Test instrument indications
instrument
Calibrator
range settings
output settings
Min
Max
320
mV
300  mV
299
301
1
3.2 V
3V
2.99
3.01
32
V
30  V
29.9
30.1
320
V
300  V
299
301
1000
V
900  V
895
905
1
Set function switch to V dc.
Table 10b. Model 77 Series III Ac Voltage
Performance Checks
Calibrator output
settings
Test instrument indications
Test
instrument
Voltage
Frequency
range settings
(V)
(Hz)
Min
Max
3.2 V
3
45
2.938
3.062
3.2 V
3
500
2.938
3.062
32
V
30
45
29.38
30.62
32
V
30
1000
29.38
30.62
320
V
300
1000
293.8
306.2
320
V
300
45
293.8
306.2
1000
V
900
45
880
920
1000
V
900
1000
880
920
Table 10c. Model 77 Series III Resistance
Performance Checks
Calibrator
Test instrument
Error
Output
range settings
display indications
settings
( )
(%)
( )
1
320
100
0.8
3200
1k
0.6
32
k
10  k
0.6
2
320
k
100  k
0.6
3.2  M
1M
0.6
32
M
10  M
2.1
1
Set calibrator 2 wire Comp ON.
2
Set calibrator 2 wire Comp OFF.
Table 10d. Model 77 Series III Dc Current
Performance Checks
Test instrument
Calibrator
Test instrument indications
range settings
output settings
Min
Max
32  mA
30 mA
29.53
30.47
320  mA
300  mA
295.3
304.7
1
10  A
5A
4.9
5.1
1
Connect calibrator amplifier CURRENT OUTPUT HI and LO to TI 10A and COM.