TB 9-6625-2243-35
a. Performance Check
(1) Set function switch to 300 mV dc (mV dc on some models).
(2) Set TI and calibrator to settings listed in table a for TI model Dc Voltage. TI will
indicate within limits specified.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
a. Performance Check
(1) Set function switch to V ac.
(2) Set TI and calibrator to settings listed in table b for TI model Ac Voltage. TI will
indicate within limits specified.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
a. Performance Check
(1) Set function switch to Ω.
(2) Set TI and calibrator to settings listed in table c for TI model Resistance. At
each setting, use calibrator output adjustment controls to set calibrator control display
Reading equal to TI indication. Calibrator control display Error indications will be within
limits specified.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.
a. Performance Check
(1) For model 73 only, connect calibrator amplifier CURRENT OUTPUT HI and
LO to TI 10A and COM.
(2) For all other models, connect calibrator OUTPUT HI and LO to TI 300 mA and COM.
(3) Set function switch to A dc.
(4) Set TI and calibrator to settings listed in table d for TI model Dc Current. TI
will indicate within limits specified.
b. Adjustments. Refer to TI manufacturer's manual for corrective action.