SECTION III. CALIBRATION PROCESS
Dc Voltage

Calibration Procedure For Multimeter Model 3458A Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
TB 9-6625-2246-40
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(s)  to  minimum  after  each  step  within  the
performance check where applicable.
a. Press Power On/Off pushbutton to On and allow at least 4 hours for equipment to
warm-up and stabilize.
b. Release Terminals Front/Rear pushbutton to Front.
c. Press keys as listed in (1) through (3) below:
(1) FUNCTION/RANGE blue shift then NUMERIC/USER Menu/E.
(2) FUNCTION/RANGE Menu Scroll  until MENU FULL is displayed.
(3) NUMERIC/USER Enter.
d. Press keys as listed in (1) through (3) below:
(1) FUNCTION/RANGE blue shift then MENU T/Recall State.
(2) FUNCTION/RANGE Menu Scroll  until TEMP? is displayed.
(3) NUMERIC/USER Enter.
e. Record displayed temperature as current internal temperature.
f. Press keys as listed in (1) through (3) below:
(1) FUNCTION/RANGE blue shift then MENU C/Auto Cal.
(2) FUNCTION/RANGE Menu Scroll  until CAL? is displayed.
(3) NUMERIC/USER 58 then Enter.
g. Current internal temperature of TI recorded in e above must be within 5 degrees
Celsius of its displayed calibration temperature when last adjusted.
h. Repeat technique of f and g above for CAL? 59 and CAL? 60.
i. Verify all inputs are disconnected and press FUNCTION/RANGE blue shift key
then Test/  key.
NOTE
Once the test is completed, the display shows SELF TEST
PASSED or SELF TEST FAILED.  If SELF TEST FAILED
take corrective action before continuing.