TB 9-6625-2248-24
(13) Set function generator for a sine wave, 10 Hz, 1 mVp-p, 50 output. Increase
function generator amplitude until TI displays a stable indication. Function generator
amplitude will not exceed 28 mVp-p.
(14) Repeat (13) above for frequencies from 10 Hz to 10 MHz.
(15) Release CHANNEL A and CHANNEL B AC/DC pushbuttons to AC.
(16) Repeat (13) above for frequencies from 30 Hz to 10 MHz.
(17) Press CHANNEL A and CHANNEL B AC/DC pushbuttons to DC.
(18) Disconnect function generator from TI.
(19) Connect signal generator RF OUTPUT to TI INPUT A using 50
feedthrough
termination.
(20) Set signal generator amplitude to 25 mV and adjust frequency from 10 to 100
MHz. TI will display a stable indication.
b. Adjustments. No adjustments can be made.
9. Sensitivity Channel C (Option 003)
a. Performance Check
(1) Press FREQ C pushbutton and adjust GATE TIME/DELAY control to MIN.
(2) Connect signal generator RF OUTPUT to TI INPUT C.
(3) Set signal generator to the frequencies listed in table 3. Adjust signal generator
amplitude until TI displays a stable indication of the applied frequency. Adjusted signal
generator amplitude will not exceed limits specified.
Adjusted signal
Signal generator output
generator
(MHz)
amplitude
50
15 mV
150
15 mV
350
15 mV
650
15 mV
700
75 mV
900
75 mV
1000
75 mV
b. Adjustments. No adjustments can be made.
a. Performance Check
NOTE
Do not perform power supply checks if all other parameters are
within tolerance.