Displayed Average Noise Level Test
Resolution Bandwidth Accuracy and Selectivity Test

Calibration Procedure For Spectrum Analyzer An/Usm-489A Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
TB 9-6625-2250-24
(7) Press TI FREQUENCY key then [CENTER FREQ] to 99 kHz.
(8) Press TI CONTROL TRIG key then [SINGLE] key. TI MKR amplitude
indication will be <-100 dBm.
(9) Press TI keys and enter values using DATA keys as listed in (a) through (e) below:
FREQUENCY and then [START FREQ] to 1 MHz.
(a)
[STOP FREQ] to 2.9 GHz.
(b)
MARKER OFF.
(c)
CONTROL BW then [RES BW] to 1 MHz.
(d)
[VIDEO BW] to 10 kHz.
(e)
(10) Press TI keys as listed in (a) through (d) below:
CONTROL TRIG.
(a)
[SINGLE] wait for completion of sweep.
(b)
MARKER ON.
(c)
[MKRNOISE ON].
(d)
(11) Adjust TI knob to position marker at the highest average noise level on TI crt.
(12) Press TI keys and enter values using DATA keys as listed in (a) through (h) below:
MARKER MKR  then [MARKER  CF].
(a)
SPAN then [ZERO SPAN].
(b)
MARKER OFF.
(c)
CONTROL BW then [RES BW] to 100 Hz.
(d)
[VIDEO BW] to 1 Hz.
(e)
CONTROL TRIG.
(f)
[SINGLE] wait for completion of sweep.
(g)
MARKER ON.
(h)
(13) TI MKR amplitude indication will be <-121 dBm.
(14) Press TI keys and enter values using DATA keys as listed in (a) through (e) below:
(a) FREQUENCY then [START FREQ] to 2.9 GHz.
(b) [STOP FREQ] to 6.46 GHz.
(c) MARKER OFF.
(d) CONTROL BW then [RES BW] to 1 MHz.
(e) [VIDEO BW] to 10 kHz.
(15) Repeat (10) through (12) above. TI MKR amplitude indication will be <-121 dBm.
(16) Press TI keys and enter values using DATA keys as listed in (a) through (e) below:
(a) FREQUENCY then [START FREQ] to 6.46 GHz.
(b) [STOP FREQ] to 13 GHz.
(c) MARKER OFF.
9