TB 9-6625-2250-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
MEASURING RECEIVER
Frequency range: 300 MHz
Measuring receiver system N5530S
Accuracy: 300 Hz
consisting of: Spectrum Analyzer,
dBm range: -9.7 to -10.3
Agilent Model E4440A (E4440A),
Accuracy: 0.075 dBm
Power meter, Agilent Model E4419B
Frequency response: 10 MHz to 18
(E4419B), and Sensor module, Agilent
GHz (referenced to 300 MHz) at 10
Model N5532A opt. 504 (504), 518
dBm
(518), 526 (526)
1
Accuracy:
Range: 1000 Vdc
Fluke, Model 8840A/AF-05 (AN/GSM-
Accuracy: 2 Vdc
64D)
Weinschel, Model 1870A (7916839)
POWER SPLITTER
Frequency range: 10 MHz to 18 GHz
Power range: -10 dBm
Accuracy:1
SYNTHESIZED SIGNAL
Anritsu, Model 68369NV (68369NV)
Power range: 0 to -15 dBm
Accuracy: 1.525 dBm
Hewlett-Packard, Model
SYNTHESIZER/LEVEL
3335AOPT001-KO6 (MIS-35938)
Amplitude range: -80 to +10 dBm
Accuracy: 0.02 dB per 10 dB step
Combined accuracy of measuring receiver and power splitter is 1.525 dBm
1
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Additional maintenance information is contained in TM 11-6625-3250-40.
d. When indications specified in paragraphs 8 through 20 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 8 through 20. Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to TI.