TB 9-6625-2250-35
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Marker amplitude accuracy
Measured at 300 MHz
Reference levels (dBm)
Accuracy (<dB)
0
3.3
-10
3.3
-40
3.3
-50
3.3
1Limited
to 18.0 GHz due to N type input connector.
2N
3Procedure limits of 5 GHz
EQUIPMENT REQUIREMENTS
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-287. Alternate items may be used by the calibrating activity. The
items selected must be verified to perform satisfactorily prior to use and must bear evidence
of current calibration. The equipment must meet or exceed the minimum use specifications
listed in table 2. The accuracies listed in table 2 provide a four-to-one ratio between the
standard and TI.
5. Accessories Required. The
accessories required for this
calibration are common
usage accessories issued as indicated in 4 above and are not listed in this calibration
procedure.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
Weinschel, Model 9918-6 dB (9918-6
ATTENUATOR (FIXED)
6 dB:
dB), Model 9918-10 dB (9918-10 dB)
and Model 9918-20dB (9918-20 dB)
Accuracy: 0.3 dB
10 dB:
Accuracy: 0.5 dB
20 dB:
Accuracy: 0.5 dB
ATTENUATOR NO. 1
Range: 0 to 12 dB
Hewlett-Packard, Model 355C
Frequency: 50 MHz
(7910807)
Hewlett-Packard, Model 355D (355D)
ATTENUATOR NO. 2
Range: 0 to 60 dB
Frequency: 300 MHz
Accuracy with correction report:
0.825 dB