TB 9-6625-2250-35
(16) TI MKR indication will be within 6.1 dB of indication recorded in (6) above.
Record TI MKR indication.
(17) Disconnect signal generator from TI.
(18) Connect TI 10 MHz REF IN/OUT to synthesizer/level generator 40/N MHz
REF INPUT.
(19) Connect synthesizer/level generator OUTPUT 50Ω to TI INPUT 50Ω.
(20) Press synthesizer/level generator keys to values as listed in (a) through (c)
below:
(a) FREQUENCY to 10 MHz.
(b) AMPLITUDE to -4 dBm.
(c) AMPTD INCR to 0.1 dB.
(21) Press TI MARKER PEAK SEARCH key.
(22) Slowly adjust synthesizer/level generator amplitude until TI MKR indication is
as close as possible to the indication recorded in (16) above.
(24) Press FREQUENCY key then [CENTER FREQ] to 1 MHz.
(25) Press MARKER PEAK SEARCH key.
(26) TI MKR indication will be within 6.1 dB of indication recorded in (6) above.
[CENTER FREQ] 100 kHz and 10 kHz settings.
NOTE
tolerance.
a. Performance Check
NOTE
Perform measuring receiver and sensor module (11792A) ZERO
and CALIBRATE.
36 CHANGE 1
PIN: 069371-001