TB 9-6625-2253-35
(3) Rotate calibrator knob below EDIT FIELD pushbutton to adjust calibrator
display indication to equal TI indication. Calibrator Err display indication will be within
limits specified in first row of table 6.
(4) Repeat technique of (2) and (3) above, using the settings and indications listed in
table 6. Calibrator Err display indication will be within
limits specified in
table 6.
Test instrument
Calibrator
Range
Output
Err indication
(%)
200
190.0 Ω
0.37
Ω
2 kΩ
1.9 kΩ
0.16
20 kΩ
19
kΩ
0.16
1
200 kΩ
190
kΩ
0.16
2000 kΩ
1.9 MΩ
0.21
20 MΩ
10 MΩ
2.1
10 MΩ
200 nS
3.0
2-wire comp: OFF
1
b. Adjustments. No adjustments can be made.
a. Deenergize and disconnect all equipment.
b. Annotate and affix DA label/form in accordance with TB 750-25.
CHANGE 1 7/(8 Blank)