TB 9-6625-2254-24
IDENTIFICATION AND DESCRIPTION
This bulletin provides instructions for the
1. Test Instrument Identification.
calibration of Digital Multimeter, Tektronix, Types DM 511, DM 5110, and DM 511 Opt 2.
The manufacturer's manual was used as the prime data source in compiling these
instructions. The equipment being calibrated will be referred to as the TI (test instrument)
throughout this bulletin.
a. Model Variations. Model DM 5110 is programmable using IEEE-488 interface.
Option 2 includes a Tektronix P6602 temperature probe.
b. Time and Technique. The time required for this calibration is approximately 2
hours, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. When adjustments are in tables, the (R) follows the designated adjustment.
Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications
parameters
Accuracy
Dc voltage
Range
(0.05% of reading + 0.015% FS)
200 mV
2V
(0.05% of reading + 0.01% FS)
20 V
(0.05% of reading + 0.015% FS)
200 V
(0.05% of reading + 0.01% FS)
1000 V
(0.05% of reading + 0.02% FS)
Dc current
Range: 0 to 2000 mA in 5 steps
Accuracy: (0.1% of reading +0.01% FS)
Ac voltage
Range: 0 to 500 V, 20 Hz to 50 kHz in 5 steps
Accuracy: 20 to 100 Hz (0.8% of reading + 0.05% FS)
100 Hz to 10 kHz (0.3% of reading + 0.05% FS)
10 to 20 kHz (0.6% of reading + 0.05% FS)
20 to 50 kHz (1.0% of reading + 0.05% FS)
Ac current
Range: 0 to 2000 mA in 5 steps
Accuracy: 20 to 100 Hz (0.8% of reading +0.05% FS)
100 to 10 kHz (0.3% of reading +0.05% FS)
Ac dBV
Range: -60 to +54 dBV
Accuracy: -34 to +54 dBV .0.3 dB 20 Hz to 20 kHz
-54 to -34 dBV 0.6 dB 20 Hz to 10 kHz
-60 to -54 dBV 1.0 dB 20 Hz to 10 kHz