TB 9-6625-2262-24
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification.
This bulletin provides instructions for the
calibration of Multimeter Fluke, Model 8025A and APN 13235191 (Fluke, Model 8025B).
The manufacturer's manual and specification control drawing No. 13235191 were used as
the prime data sources in compiling these instructions. The equipment being calibrated
will be referred to as the TI (test instrument) throughout this bulletin.
a. Model Variations. Variations among models are described in text.
b. Time and Technique. The time required for this calibration is approximately 1
hour, using the dc and low frequency technique.
2. Forms, Records, and Reports
a. Forms, records, and reports required for calibration personnel at all levels are
prescribed by TB 750-25.
b. Adjustments to be recorded are designated (R) at the end of the sentence in which
they appear. Report only those adjustments made and designated with (R).
3. Calibration Description. TI parameters and performance specifications which
pertain to this calibration are listed in table 1.
Test instrument
Performance specifications
(% of reading + digits) (31/2 digit display)
parameters
Dc voltage
Range: 0 to 1000 V
Accuracy: 0.2 + 1
Ac voltage
Accuracy
Range
40 Hz to 2 kHz
2 to 10 kHz
10 to 30 kHz
320.0 mV
0.5 + 3
2.0 + 3
4.0 + 10
3.200 V
32.00 V
320.0 V
1000 V
1.0 + 3
3.0 + 3
N/A
Range
Accuracy
0.3 + 2
320.0
33.200 k
.2 + 1
3 2.00 k
23.0 k
0
0
.200 M
1.0 + 1
32.00 M
32.00 nS
2.0 + 10
Dc current
Range: 0 to 10 A
Accuracy: 0.75 + 2
Ac current
Range: 0 to 10 A
Accuracy: 1.5 + 2