TB 9-6625-2263-24
Table 2. Minimum Specifications of Equipment Required - Continued
Minimum use
Manufacturer and model
Common name
specifications
(part number)
CALIBRATOR
Ac voltage:
(continued)
Range
Accuracy
(%)
1
V
15 Hz
1.18
1
V
30 Hz
0.45
1
V
200 Hz
0.0625
1
V
15 kHz
0.0625
1
V
90 kHz
0.8
1
V
200 kHz
1.5
1
V
1 MHz
6.25
10
V
15 Hz
1.12
10
V
30 Hz
0.39
10
V
200 Hz
0.058
10
V
15 kHz
0.058
10
V
90 kHz
0.74
10
V
200 kHz
1.4
10
V
1 MHz
5.75
50
V
200 kHz
1.65
100
V
15 Hz
1.07
100
V
30 Hz
0.345
100
V
200 Hz
0.538
100
V
15 kHz
0.538
100
V
90 kHz
0.718
600
V
200 Hz
0.059
600
V
15 kHz
0.059
600
V
30 kHz
0.76
Range: 4.9 to 5.2 V dc
Fluke, Model 8840A/AF05
Accuracy: 0.74%
(AN/GSM-64D)
A
Beckman, Model CR10M
Range: 10 M
ccuracy: 0.0275%
STANDARD
(8598965)
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in the manufacturer's manual for this TI.