TB 9-6625-2264-35
5. Accessories Required. The
accessories required for this
calibration are
common
usage accessories issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessories, supplied with the standard,
are also required for this calibration: Bit, tamper proof, 1/8 inch, PN MISC 003-F27503,
manufacturer: Corland; and lens cap/target, APN 13440044.
Table 2. Minimum Specifications of
Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
NIGHT VISION DEVICE
Wavelength range: 305 to 1000 nm
13335470 (13335470)
DETECTOR STANDARD
wavelength peak response: 720 nm +50
Responsitivity: 4.73 x 107
V/W.cm-2.Sr-1at 820 nm+10%
Dynamic range: 1.0 x 10-12 to 5.0 x 10-8
w/cm2.Sr at 820 nm
Gain (elect): 109 V/A
Range: 0.82 mV to 1.71 V dc
Hewlett-Packard, Model 3458A
Resolution: 0.1 mV
(3458A)
6. Preliminary
Instructions
a. The
instructions outlined in
paragraphs 6 and
7 are preparatory to the
calibration process. Personnel should become familiar with the entire bulletin before
beginning calibration.
b. Items of equipment used in this procedure are referenced within the text by
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-5855-264-14 for this TI.
d. Unless otherwise specified, all controls and control settings refer to the TI.
WARNING
HIGH VOLTAGE is used or exposed during the performance
of this calibration. DEATH ON CONTACT may result if
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove protective cover from TI as necessary to gain access to adjustments.