TB 9-6625-2267-35
WARNING
HIGH VOLTAGE is used or exposed during the performance
personnel fail to observe safety precautions. REDUCE
OUTPUT(S) to minimum after each step within the
performance check where applicable.
a. Remove protective cover from TI only to make adjustments and replace upon
completion.
b. Connect TI to a 115 V ac source.
c. Press POWER ON/OFF pushbutton to ON and allow at least 15 minutes for
stabilization.
a. Performance Check
(1) Connect calibrator OUTPUT HI and LO to TI INPUT V-Ω and LO.
(2) Press TI FUNCTION DCV pushbutton.
(3) Press TI RANGE pushbuttons and set calibrator output for settings listed in
table 3. If TI
indications are not within limits
specified, perform b below.
Test instrument
Test instrument
RANGE
Calibrator
indications
pushbuttons
output settings
Min
Max
200 mV
190 mV
+189.90
+190.10
200 mV
-190 mV
-189.90
-190.10
2
1.9 V
+1.8995
+1.9005
2
-1.9 V
-1.8995
-1.9005
20
19 V
+18.995
+19.005
20
-19 V
-18.995
-19.005
200
190 V
+189.95
+190.05
1200 V
1000 V
+999.7
+1000.3
b. Adjustments
(1) Set calibrator for a 1 MΩ output.
(2) Press TI RANGE 200mV pushbutton.
(3) Adjust C24 (fig. 1) until TI indicates between -00.01 and +00.01.
(4) Set calibrator for a 190 mV dc output. Record TI indication.