SECTION IV. CALIBRATION PROCESS FOR FLUKE, MODEL 80J-10
Table 3. Shunt Accuracy

Calibration Procedure For Ac Divider, Model 7405A-4207, And Current Shunt, Model 80J-10 Page Navigation
  1    2    3    4  5  6  
img
. . . . . . .
TB 9-6625-2275-24
Figure 2. Shunt accuracy.
12. Shunt Accuracy
a. Performance Check
(1) Set calibrator for a 1 A dc output. Multimeter will indicate within the limits
specified in first row of table 3.
(2) Set calibrator to STANDBY and connect equipment as shown in figure 2
CONNECTION B.
(3) Repeat technique of (1) above for calibrator outputs listed in table 3.
Multimeter will indicate within the limits specified in table 3.
b. Adjustments. No adjustments can be made.