TB 9-6625-2277-24
(10) Repeat (1) and (3) through (9) above until no further adjustments are required.
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI mA and calibrator OUTPUT LO to
TI COMMON.
(2) Press mA/A (mA on models 8012A and 8010M) function and AC/DC pushbutton
to DC.
(3) Press TI range pushbutton and set calibrator output for settings listed in table 5.
TI will indicate within limits specified in table 5.
Test instrument
Test instrument
range
Calibrator
indications
pushbuttons
settings
Min
Max
A
A
189.3
190.7
200
190
2 mA
1.9 mA
1.893
1.907
20 mA
19 mA
18.93
19.07
200 mA
190 mA
189.3
190.7
2000 mA
1.9 A
1893
1907
10 Al
10 A
9.94
10.06
For models 8010A and 8010A-01 only, connect amplifier output to TI 10A and
l
COMMON.
b. Adjustments. No adjustments can be made.
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI V/k /S (V/k
on model 8010M) and
calibrator OUTPUT LO to TI COMMON.
(2) Press k /S function pushbutton.
(3) Press TI range pushbutton and set calibrator to the nominal resistance outputs
as listed in table 6. At each resistance output, rotate calibrator knob below EDIT FIELD
pushbutton to adjust calibrator display indication to equal TI indication. The calibrator err
display indication will be within limits specified in table 6.
Test
Calibrator
err
instrument
Calibrator
display indications
range
output
(%)
pushbuttons
nominal value
0.3
200
100
0.3
2
1k
0.3
20
10 k
0.3
200
100 k
0.6 (0.3)1
2000
1M
0.6
20 M
10 M