TB 9-6625-2277-24
(10) Repeat (1) and (3) through (9) above until no further adjustments are required.
10. Dc Current Accuracy
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI mA and calibrator OUTPUT LO to
TI COMMON.
(2) Press mA/A (mA on models 8012A and 8010M) function and AC/DC pushbutton
to DC.
(3) Press TI range pushbutton and set calibrator output for settings listed in table 5.
TI will indicate within limits specified in table 5.
Table 5. Dc Current
Test instrument
Test instrument
range
Calibrator
indications
pushbuttons
settings
Min
Max
A
A
189.3
190.7
200
190
2  mA
1.9  mA
1.893
1.907
20  mA
19  mA
18.93
19.07
200  mA
190  mA
189.3
190.7
2000  mA
1.9  A
1893
1907
10  Al
10  A
9.94
10.06
For models 8010A and 8010A-01 only, connect amplifier output to TI 10A and
l
COMMON.
b. Adjustments. No adjustments can be made.
11. Resistance
a. Performance Check
(1) Connect calibrator OUTPUT HI to TI V/k /S (V/k
on model 8010M) and
calibrator OUTPUT LO to TI COMMON.
(2) Press k /S function pushbutton.
(3) Press TI range pushbutton and set calibrator to the nominal resistance outputs
as listed in table 6. At each resistance output, rotate calibrator knob below EDIT FIELD
pushbutton to adjust calibrator display indication to equal TI indication. The calibrator err
display indication will be within limits specified in table 6.
Table 6. Resistance
Test
Calibrator
err
instrument
Calibrator
display indications
range
output
(%)
pushbuttons
nominal value
0.3
200
100
0.3
2
1k
0.3
20
10 k
0.3
200
100 k
0.6 (0.3)1
2000
1M
0.6
20  M
10 M
8